A sensor composed of an indicator and a
reference electrode in contact with a
thin film of solution which is separated from the bulk of the sample solution by a gas-permeable
membrane or an air gap. This intermediate solution interacts with the gaseous species (penetrated
through the
membrane or an air gap) in such a way as to produce a change in a measured
constituent (e.g. the
H+
activity) of the intermediate solution. This change is then sensed by the
ion-selective electrode and is related to the
partial pressure of the gaseous species in the sample. [Note: In electrochemical literature the term
gas electrode is used for the classical, redox-equilibrium-based gas electrodes as
well, such as the hydrogen or the chlorine gas electrodes (
Pt
(s)|
H2
(g) |
H+
(aq) or
Pt
(s) |
Cl2
(g) |
Cl−
(aq)]. These electrodes respond both to the
partial pressure of the gas (
H2
or
Cl2
) and to the ionic activities (
H+
or
Cl−). The Clark oxygen electrode fits under this classification although, in contrast
to other gas sensors, it is an amperometric and not a potentiometric device.